Skip to Content
Digital View
  • Home
  • Products
    • Controller Boards
    • E Ink Solutions
    • Accessories
    • AI @ Display Edge
    • Services
  • CSG
  • Applications
    • AI Enabled Displays
    • Kiosks
    • Embedded Systems
    • Marine
    • Digital Signage Displays
    • Industrial Displays
    • Defense Related Displays
    • Transport Displays
    • Medical Displays
  • How to Buy
    • Resellers
    • Contact Sales
  • Support
    • Compliance
    • Contact Support
  • About
  • Blog
  • 0
  • Contact
Digital View
  • 0
    • Home
    • Products
      • Controller Boards
      • E Ink Solutions
      • Accessories
      • AI @ Display Edge
      • Services
    • CSG
    • Applications
      • AI Enabled Displays
      • Kiosks
      • Embedded Systems
      • Marine
      • Digital Signage Displays
      • Industrial Displays
      • Defense Related Displays
      • Transport Displays
      • Medical Displays
    • How to Buy
      • Resellers
      • Contact Sales
    • Support
      • Compliance
      • Contact Support
    • About
    • Blog
  • Contact

DICOM and SVX-4096

  • Digital View Blog
  • DICOM and SVX-4096
  • August 19, 2025 by
    DICOM and SVX-4096
    James Henry

    DICOM Support with the SVX-4096 LCD Controller


    Digital View’s SVX-4096 LCD controller for 4K resolution LCD panels includes support for DICOM lookup tables (LUTs), enabling calibration to the DICOM grayscale standard display function (GSDF).

    What DICOM Calibration Means for Displays

    DICOM (Digital Imaging and Communications in Medicine) is a widely used standard for handling and viewing medical images. Part 14 of the standard defines the GSDF, which is specifically concerned with how grayscale tones are displayed. Although this standard addresses grayscale response, it is applied in the context of full-color displays to ensure that brightness and contrast are rendered consistently for medical imaging.

    How the SVX-4096 Supports DICOM LUTs

    With the SVX-4096, integrators can:

    • Use a color analyzer to measure a particular LCD panel’s response.
    • Generate a gamma curve LUT based on the DICOM GSDF.
    • Upload this LUT into the SVX-4096 firmware via the DDC-CI channel.
    • Select the DICOM LUT from the controller’s OSD settings.

    Because different LCD panels vary in characteristics, calibration is panel-specific. Many current 4K LCD panels, however, are suitable for this process.

    Technical Notes

    • The SVX-4096 uses a 10-bit gamma LUT, requiring 6144 bytes per curve.
    • Up to eight LUTs can be stored in firmware, supporting multiple calibration options.
    • LUT upload is performed through the DDC-CI channel, allowing connection to a PC tool during calibration .
    • Optional DDC commands support writing brightness parameters and restoring factory defaults.
    • See the Downloads section of the SVX-4096 product page for an application note.

    Typical Uses

    Adding a DICOM LUT makes the SVX-4096 suitable for display systems where compliance with DICOM calibration is required, such as:

    • Radiology review and clinical workstations.
    • PACS (Picture Archiving and Communication System) displays.
    • Any setting where consistent rendering of medical images is needed.

    Summary

    The SVX-4096 combines high-resolution 4K display control with the ability to support DICOM calibration. By providing a path to upload and manage DICOM LUTs, it gives manufacturers and integrators the tools to develop medical display solutions that meet DICOM grayscale requirements while retaining the full capabilities of a color LCD controller.


    Next Steps

    • See the SVX-4096 product page
    • Browse our complete range of controller boards
    • Contact us

    Composite Video Revisited: AVD-2000

    Previous blog - pre May 2025

    Digital View - enabling digital displays

    Digital View — Engineering display electronics since 1995. We design and manufacture highly reliable controllers for LCD, OLED, and E Ink ePaper displays, powering mission-critical applications in defense, medical, industrial, and digital signage. 

    Main Pages

    Contact Us
    Controller Summary
    Office Addresses
    Privacy Policy
    About

    Phone

    USA +1-408 782 7773 (Morgan Hill, California)
    UK +44(0)20 7631 2150 (London)
    HK +852-2861 3615 (Hong Kong)

    Follow us
    ​ ​

      Contact Us: 

      Sales inquiries or any other questions...

    Submit
    Copyright © 2025 Digital View
    Digital View - Enabling Displays

    We use cookies to provide you a better user experience on this website. Cookie Policy

    Only essentials I agree